Description
Design
Overall diameter …………… 11 mm
Optic diameter …………… 6 mm
Platform ………………….. Monolithic, 4-point fixation, square edge 360 degrees
Optic design …………….. Aspherical design with a combination of 4th and 6th order spherical aberrations with the opposite sign
Haptic angulation ……… 0 degrees
Light filter ……….. UV filter and violet filter
Diopter series
Diopter range ……. Available cylinder values
+6.0 D ………………………… +0.75 D / +1.0 D / +1.5 D
+7.0 D ………………………… +0.75 D / +1.0 D / +1.5 D / +2.25 D / +3.0 D / +3.75 D / +4.5 D
+8.0 D; +9.0 D; +10.0 D … +0.75 D / +1.0 D / +1.5 D / +2.25 D / +3.0 D / +3.75 D / +4.5 D / +5.25 D / +6.0 D
from +10.5 D to +30.0 D
(step 0.5 D) ………………….. +0.75 D / +1.0 D / +1.5 D / +2.25 D / +3.0 D / +3.75 D / +4.5 D / +5.25 D / +6.0 D
+31.0 D; +32.0 D ………….. +0.75 D / +1.0 D / +1.5 D / +2.25 D / +3.0 D / +3.75 D / +4.5 D / +5.25 D / +6.0 D
+33.0 D ………………………. +2.25 D / +3.0 D / +3.75 D / +4.5 D / +5.25 D / +6.0 D
+34.0 D ………………………. +4.5 D / +5.25 D / +6.0 D
Constants
SRK/T A constant ultrasound biometry ……………… 118.8
SRK/T A constant optical biometry … 118.5
Hoffer Q pACD ………………………………………. 5.23
Holladay | Surgeon Factor ………………………. 1.48
EVO A constant …………………………………….. 118.5
Barrett A constant ………………………………….. 118.4
Lens Factor: 1.57
Hill-RBF A constant …………………………….. 118.32
Implantation System
IOL pre-installed in the injector …………… Accuject Pro
Recommended incision size ……………… 2.2 mm
Instructions for use can be downloaded here